| ARCHETYPE ID | openEHR-EHR-CLUSTER.ventilator_settings2.v0 |
| Concept | Ventilator settings/findings |
| Description | Details of ventilator settings and reported findings |
| Purpose | To record details of ventilator settings, such as ventilation type, volume, frequencies and pressures. |
| References | Derived from: <Add reference to original resource here> |
| Copyright | © openEHR Foundation, HiGHmed |
| Authors | Author name: Dr Ian McNicoll Organisation: Ocean Informatics, United Kingdom Email: ian.mcnicoll@oceaninformatics.com Date originally authored: 25/10/2010
|
| Other Details Language | Author name: Dr Ian McNicoll Organisation: Ocean Informatics, United Kingdom Email: ian.mcnicoll@oceaninformatics.com Date originally authored: 25/10/2010
|
| Other Details (Language Independent) | - Licence: This work is licensed under the Creative Commons Attribution-ShareAlike 3.0 License. To view a copy of this license, visit http://creativecommons.org/licenses/by-sa/3.0/.
- Custodian Organisation: HiGHmed
- References: Derived from:
- Original Namespace: org.highmed
- Original Publisher: HiGHmed
- Custodian Namespace: org.highmed
- MD5-CAM-1.0.1: EBA1A7732F98B4B0DD5CBBCA7633F7E9
- Build Uid: 4e5a5d75-d2df-465f-b659-bfa41418aee9
- Revision: 0.0.1-alpha
|
| Keywords | ventilator, assisted |
| Lifecycle | in_development |
| UID | c580786d-cac7-4dda-8b09-a874e908e413 |
| Language used | en |
| Citeable Identifier | 1246.145.1130 |
| Revision Number | 0.0.1-alpha |
| items | |
| Type of ventilation | Type of ventilation: Type of ventilation of patient
- at0145::INVASIVE [Invasive type of ventilation]
- at0146::NON-INVASIVE [Non-invasive type of ventilation]
|
| Ventilation mode | Ventilation mode: The mode of mechanical ventilation used.
- at0104::A/C [Assist-control ventilation]
- at0108::SIMV [Synchronised intermittent mandatory ventilation.]
- at0105::BILEVEL [BILEVEL]
- at0065::SPONTANEOUS [Patient is breathing spontaneously.
]
- at0106::PS [Pressure support ]
- at0077::MMV [Mandatory minute ventilation.
]
- at0152::SIMV / ASB [SIMV / ASB]
- at0076::IPPV [Intermittent positive pressure ventilation.
]
- at0070::CPAP [Pressure support / Continous positive airway pressure.
]
- at0072::CMV [Controlled mandatory ventilation ]
- at0079::BIPAP [Biphasic positive airway pressure.
]
- at0153::BIPAP / ASB [BIPAP / ASB]
- at0073::HFO [ High frequency oscillation ventilation.
]
- at0075::PTV [Patient triggered ventilation.
]
- at0080::PSV [Pressure support ventilation.
]
- at0083::PCV [Pressure controlled ventilation.
]
- at0084::S/T [Spontaneous timed breathing.
]
- at0085::PAV/T [Proportional assist ventilation/Time.
]
- at0097::Sigh [Sigh.]
- at0155::HFO + CMV [HFO + CMV mode]
|
| Ventilation submode | Ventilation submode: Ventilation submode
- at0151::TC [TC]
- at0147::VS [VS]
- at0124::PS [Pressure support]
- at0143::VS+ [VS+]
- at0117::PC/PS [PC/PS]
- at0118::PC/TC [ PC/TC]
- at0119::VC/PS [VC/PS]
- at0120::VC/TC [VC/TC]
- at0129::CV [CV]
- at0131::CMV [Controlled mechanical ventilation.]
- at0116::VC [Volume controlled]
- at0122::TS [TS]
- at0115::PC [Pressure controlled.]
- at0135::ASB [Assisted spontaneous breathing.]
- at0138::Assist [Assist]
- at0139::ST [ST]
- at0140::PS/CPAP [ PS/CPAP]
- at0141::CPAP/APRV [CPAP/APRV]
- at0142::ASB/Assist [ASB/Assist]
|
| Ventilation device | Ventilation device: Details of the ventilation device. Include: openEHR-EHR-CLUSTER.device.v1 and specialisations |
| Oxygen delivery | Oxygen delivery: Details of oxygen delivery. Include: openEHR-EHR-CLUSTER.gas_delivery-oxygen.v1 and specialisations |
| Heater used | Heater used: If true a heater should be used/ is used. |
| (Frequency) | (Frequency): A ventilator frequency setting. Property: Frequency Units: - 0.0..500.0 /min
Limit decimal places: 0 - 0.0..500.0 Hz
Limit decimal places: 0
Runtime name constraint:
- at0010::Respiratory rate (f) [Respiratory rate, frequency or BPM breaths per minute.]
- at0111::Total respiration rate (f tot) [Total respiration rate]
- at0008::Min respiratory rate [Minimum respiratory rate.
]
- at0009::Safe respiratory rate [Safe respiratory rate.
]
- at0156::Mandatory raspiration rate [Mandatory raspiration rate]
- at0161::HFV Frequency [HFV Frequency ]
|
| (Pressure) | (Pressure): A ventilator pressure setting. Property: Pressure Units: - 0.0..200.0 mbar
Limit decimal places: 1 - >=0.0 cm[H20]
Runtime name constraint:
- at0091::Peak airway pressure (P peak) [Peak airway pressure (P peak)]
- at0026::Mean airway pressure (P mean) [Mean airway pressure (P mean)]
- at0092::Plateau airway pressure (P plateau) [Plateau airway pressure (P plateau)]
- at0017::Positive End Expiratory Pressure (PEEP) [Positive End Expiratory Pressure (PEEP)]
- at0154::Assisted Spontaneous Breath Setting (PEEP asb) [Assisted Spontaneous Breath Setting (PEEP asb)]
- at0024::Maximum inspiratory pressure (P max) [Maximum inspiratory pressure (P max)]
- at0150::End Inspiratory Pressure (P i) [End Inspiratory Pressure (P i)]
- at0028::Support Pressure (PS) [Support Pressure (PS)]
- at0020::Hi (PEEP) [High level positive expiratory end pressure.]
- at0021::Lo (PEEP) [Low level positive expiratory end pressure.]
- at0025::Inspiratory pressure [inspiratory pressure]
- at0016::CPAP [Continuous positive airway pressure.]
- at0018::IPAP [Inspiratory positive airway pressure.]
- at0019::EPAP [Expiratory positive airway pressure.]
- at0027::Delta pressure [Pressure difference / delta pressure]
- at0089::Airway pressure [Airway pressure.]
- at0090::Intrinsic PEEP [Intrinsic PEEP.]
- at0096::Intermittent PEEP [Intermittent PEEP.]
- at0158::High inspiratory pressure [High inspiratory pressure ]
- at0159::HFV Mean airway pressure [HFV Mean airway pressure ]
- at0160::HFV Inspiratory pressure [HFV Inspiratory pressure ]
- at0163::HFV Amplitude [HFV Amplitude]
- at0162::Peak inspiratory pressure [Peak inspiratory pressure]
|
| (Volume) | (Volume): A ventilator volume setting. Property: Volume Units: - 0.0..2000.0 ml
Limit decimal places: 1 - 0.0..200.0 l
Limit decimal places: 1
Runtime name constraint:
- at0032::Tidal volume (V te) [Tidal volume (V te)]
- at0112::Minute volume (Ve tot) [Minute volume (Ve tot)]
- at0033::Target inspiratory volume [Target inspiratory volume.]
- at0098::Inspired tidal volume [Inspired tidal volume.]
|
| (Flow rate) | (Flow rate): A ventilator flow rate parameter. Property: Flow rate, volume Units: >=0.0 l/min Limit decimal places: 1 Runtime name constraint:
- at0149::Peak inspiratory flow (V max) [Peak inspiratory flow (V max)]
- at0099::Inspiratory flow [Inspiratory flow.]
- at0100::Expiratory flow [Expiratory flow.]
|
| (Timing) | (Timing): A ventilator duration or timing setting. Allowed values: seconds 0 seconds..1000 seconds
Runtime name constraint:
- at0039::Inspiratory time (T i) [Inspiratory time (T i)]
- at0113::Plateau time (T pl) [Plateau time (T pl)]
- at0040::Apneoa time [Apnoea time]
- at0041::Inspiratory rise time [Inspiratory rise time / inspiratory ramp]
|
| (Trigger value) | (Trigger value): The trigger value setting. Choice of:
Count min: >=0; max: <=100
Quantity Property: Flow rate, volume Units: 0.0..100.0 l/min Limit decimal places: 1
Runtime name constraint:
- at0047::Flow sensor [Flow sensor.]
- at0048::Inspiratory [Inspiratory trigger.]
- at0049::PTV sensitivity [PTV sensitivity trigger.]
- at0050::Gastric sonde [Gastric sonde trigger.]
- at0046::No trigger [A trigger is not in use.]
|
| Trigger type | Trigger type: Type of trigger applied. |
| I:E Inspiration/expiration | I:E Inspiration/expiration: Ration of inspiratory phase to expiratory phase. Numerator: 1.0..99.0 Assumed value: false
|
| I:T Inspiratory/Total | I:T Inspiratory/Total: Ratio of inspiratory phase to total phase. Numerator: 0.0..100.0
|
| Artificial airway compensation | Artificial airway compensation: Artificial airway compensation. Numerator: >=0.0 Assumed value: false
|
| NO delivered | NO delivered: Amount of Nitrogen monoxide delivered. Property: Proportion Units: 0.0..500.0 ppb (UCUM: [ppb]) Limit decimal places: 0 |
| NO2 removed | NO2 removed: Amount of Nitrogen monoxide removed. Property: Proportion Units: 0.0..5.0 ppb (UCUM: [ppb]) Limit decimal places: 0 |
| Other contributors | Biljana Prinčič, Marand d.o.o., Slovenia Uroš Rajkovič, Faculty of Organizational Sciences, University of Maribor, Slovenia |
| Translators | - German: Erik Tute, Medizinische Hochschule Hannover, reb86109@bcaoo.com, -
- Slovenian: ?
|