ARCHETYPE Ventilator settings/findings (openEHR-EHR-CLUSTER.ventilator_settings2.v0)

ARCHETYPE IDopenEHR-EHR-CLUSTER.ventilator_settings2.v0
ConceptVentilator settings/findings
DescriptionDetails of ventilator settings and reported findings
PurposeTo record details of ventilator settings, such as ventilation type, volume, frequencies and pressures.
ReferencesDerived from: <Add reference to original resource here>
Copyright© openEHR Foundation, HiGHmed
AuthorsAuthor name: Dr Ian McNicoll
Organisation: Ocean Informatics, United Kingdom
Email: ian.mcnicoll@oceaninformatics.com
Date originally authored: 25/10/2010
Other Details LanguageAuthor name: Dr Ian McNicoll
Organisation: Ocean Informatics, United Kingdom
Email: ian.mcnicoll@oceaninformatics.com
Date originally authored: 25/10/2010
Other Details (Language Independent)
  • Licence: This work is licensed under the Creative Commons Attribution-ShareAlike 3.0 License. To view a copy of this license, visit http://creativecommons.org/licenses/by-sa/3.0/.
  • Custodian Organisation: HiGHmed
  • References: Derived from:
  • Original Namespace: org.highmed
  • Original Publisher: HiGHmed
  • Custodian Namespace: org.highmed
  • MD5-CAM-1.0.1: EBA1A7732F98B4B0DD5CBBCA7633F7E9
  • Build Uid: 4e5a5d75-d2df-465f-b659-bfa41418aee9
  • Revision: 0.0.1-alpha
Keywordsventilator, assisted
Lifecyclein_development
UIDc580786d-cac7-4dda-8b09-a874e908e413
Language useden
Citeable Identifier1246.145.1130
Revision Number0.0.1-alpha
items
Type of ventilationType of ventilation: Type of ventilation of patient
  • INVASIVE [Invasive type of ventilation]
  • NON-INVASIVE [Non-invasive type of ventilation]
Ventilation modeVentilation mode: The mode of mechanical ventilation used.
  • A/C [Assist-control ventilation]
  • SIMV [Synchronised intermittent mandatory ventilation.]
  • BILEVEL [BILEVEL]
  • SPONTANEOUS [Patient is breathing spontaneously. ]
  • PS [Pressure support ]
  • MMV [Mandatory minute ventilation. ]
  • SIMV / ASB [SIMV / ASB]
  • IPPV [Intermittent positive pressure ventilation. ]
  • CPAP [Pressure support / Continous positive airway pressure. ]
  • CMV [Controlled mandatory ventilation ]
  • BIPAP [Biphasic positive airway pressure. ]
  • BIPAP / ASB [BIPAP / ASB]
  • HFO [ High frequency oscillation ventilation. ]
  • PTV [Patient triggered ventilation. ]
  • PSV [Pressure support ventilation. ]
  • PCV [Pressure controlled ventilation. ]
  • S/T [Spontaneous timed breathing. ]
  • PAV/T [Proportional assist ventilation/Time. ]
  • Sigh [Sigh.]
  • HFO + CMV [HFO + CMV mode]
Ventilation submodeVentilation submode: Ventilation submode
  • TC [TC]
  • VS [VS]
  • PS [Pressure support]
  • VS+ [VS+]
  • PC/PS [PC/PS]
  • PC/TC [ PC/TC]
  • VC/PS [VC/PS]
  • VC/TC [VC/TC]
  • CV [CV]
  • CMV [Controlled mechanical ventilation.]
  • VC [Volume controlled]
  • TS [TS]
  • PC [Pressure controlled.]
  • ASB [Assisted spontaneous breathing.]
  • Assist [Assist]
  • ST [ST]
  • PS/CPAP [ PS/CPAP]
  • CPAP/APRV [CPAP/APRV]
  • ASB/Assist [ASB/Assist]
Ventilation deviceVentilation device: Details of the ventilation device.
Include:
openEHR-EHR-CLUSTER.device.v1 and specialisations
Oxygen deliveryOxygen delivery: Details of oxygen delivery.
Include:
openEHR-EHR-CLUSTER.gas_delivery-oxygen.v1 and specialisations
Heater usedHeater used: If true a heater should be used/ is used.
(Frequency)(Frequency): A ventilator frequency setting.
Property: Frequency
Units:
  • 0.0..500.0 /min
    Limit decimal places: 0
  • 0.0..500.0 Hz
    Limit decimal places: 0

Runtime name constraint:
  • Respiratory rate (f) [Respiratory rate, frequency or BPM breaths per minute.]
  • Total respiration rate (f tot) [Total respiration rate]
  • Min respiratory rate [Minimum respiratory rate. ]
  • Safe respiratory rate [Safe respiratory rate. ]
  • Mandatory raspiration rate [Mandatory raspiration rate]
  • HFV Frequency [HFV Frequency ]
(Pressure)(Pressure): A ventilator pressure setting.
Property: Pressure
Units:
  • 0.0..200.0 mbar
    Limit decimal places: 1
  • >=0.0 cm[H20]

Runtime name constraint:
  • Peak airway pressure (P peak) [Peak airway pressure (P peak)]
  • Mean airway pressure (P mean) [Mean airway pressure (P mean)]
  • Plateau airway pressure (P plateau) [Plateau airway pressure (P plateau)]
  • Positive End Expiratory Pressure (PEEP) [Positive End Expiratory Pressure (PEEP)]
  • Assisted Spontaneous Breath Setting (PEEP asb) [Assisted Spontaneous Breath Setting (PEEP asb)]
  • Maximum inspiratory pressure (P max) [Maximum inspiratory pressure (P max)]
  • End Inspiratory Pressure (P i) [End Inspiratory Pressure (P i)]
  • Support Pressure (PS) [Support Pressure (PS)]
  • Hi (PEEP) [High level positive expiratory end pressure.]
  • Lo (PEEP) [Low level positive expiratory end pressure.]
  • Inspiratory pressure [inspiratory pressure]
  • CPAP [Continuous positive airway pressure.]
  • IPAP [Inspiratory positive airway pressure.]
  • EPAP [Expiratory positive airway pressure.]
  • Delta pressure [Pressure difference / delta pressure]
  • Airway pressure [Airway pressure.]
  • Intrinsic PEEP [Intrinsic PEEP.]
  • Intermittent PEEP [Intermittent PEEP.]
  • High inspiratory pressure [High inspiratory pressure ]
  • HFV Mean airway pressure [HFV Mean airway pressure ]
  • HFV Inspiratory pressure [HFV Inspiratory pressure ]
  • HFV Amplitude [HFV Amplitude]
  • Peak inspiratory pressure [Peak inspiratory pressure]
(Volume)(Volume): A ventilator volume setting.
Property: Volume
Units:
  • 0.0..2000.0 ml
    Limit decimal places: 1
  • 0.0..200.0 l
    Limit decimal places: 1

Runtime name constraint:
  • Tidal volume (V te) [Tidal volume (V te)]
  • Minute volume (Ve tot) [Minute volume (Ve tot)]
  • Target inspiratory volume [Target inspiratory volume.]
  • Inspired tidal volume [Inspired tidal volume.]
(Flow rate)(Flow rate): A ventilator flow rate parameter.
Property: Flow rate, volume
Units: >=0.0 l/min
Limit decimal places: 1
Runtime name constraint:
  • Peak inspiratory flow (V max) [Peak inspiratory flow (V max)]
  • Inspiratory flow [Inspiratory flow.]
  • Expiratory flow [Expiratory flow.]
(Timing)(Timing): A ventilator duration or timing setting.
Allowed values: seconds
0 seconds..1000 seconds

Runtime name constraint:
  • Inspiratory time (T i) [Inspiratory time (T i)]
  • Plateau time (T pl) [Plateau time (T pl)]
  • Apneoa time [Apnoea time]
  • Inspiratory rise time [Inspiratory rise time / inspiratory ramp]
(Trigger value)(Trigger value): The trigger value setting.
Choice of:
  •  Count
    min: >=0; max: <=100

  •  Quantity
    Property: Flow rate, volume
    Units: 0.0..100.0 l/min
    Limit decimal places: 1

Runtime name constraint:
  • Flow sensor [Flow sensor.]
  • Inspiratory [Inspiratory trigger.]
  • PTV sensitivity [PTV sensitivity trigger.]
  • Gastric sonde [Gastric sonde trigger.]
  • No trigger [A trigger is not in use.]
Trigger typeTrigger type: Type of trigger applied.
I:E Inspiration/expirationI:E Inspiration/expiration: Ration of inspiratory phase to expiratory phase.
  • Ratio
Numerator: 1.0..99.0
Assumed value: false
I:T Inspiratory/TotalI:T Inspiratory/Total: Ratio of inspiratory phase to total phase.
  • Percent
Numerator: 0.0..100.0
Artificial airway compensationArtificial airway compensation: Artificial airway compensation.
  • Percent
Numerator: >=0.0
Assumed value: false
NO deliveredNO delivered: Amount of Nitrogen monoxide delivered.
Property: Proportion
Units: 0.0..500.0 ppb
Limit decimal places: 0
NO2 removedNO2 removed: Amount of Nitrogen monoxide removed.
Property: Proportion
Units: 0.0..5.0 ppb
Limit decimal places: 0
Other contributorsBiljana Prinčič, Marand d.o.o., Slovenia
Uroš Rajkovič, Faculty of Organizational Sciences, University of Maribor, Slovenia
Translators
  • German: Erik Tute, Medizinische Hochschule Hannover, reb86109@bcaoo.com, -
  • Slovenian: ?