ARCHETYPE pure tone audiogram (openEHR-EHR-OBSERVATION.pure_tone_audiogram.v0)

ARCHETYPE IDopenEHR-EHR-OBSERVATION.pure_tone_audiogram.v0
Conceptpure tone audiogram
Descriptionpure tone audiogram
Purpose
References
AuthorsDate originally authored: 2024-02-19
Other Details LanguageDate originally authored: 2024-02-19
Other Details (Language Independent)
  • Licence: This work is licensed under the Creative Commons Attribution-ShareAlike 4.0 International License. To view a copy of this license, visit https://urldefense.com/v3/__http://creativecommons.org/licenses/by-sa/4.0/__;!!JFdNOqOXpB6UZW0!qQwlhw-IcnT2C6CvQl7W6DHGCfxFJcR5_ax9rzBY0xeyqvdD13e9TIrbz3gOjFbD-7jD6GfVRUkghy5Gh0FzH7biQaDpfyGHYQ$ .
  • Custodian Organisation: openEHR Foundation
  • Original Namespace: org.openehr
  • Original Publisher: openEHR Foundation
  • Custodian Namespace: org.openehr
  • MD5-CAM-1.0.1: 421A868385F764E2306EE0B61B20605E
  • Build Uid: 53f2e539-911a-4f01-8156-6171b572cf73
  • Revision: 0.0.1-alpha
Keywords
Lifecyclein_development
UID9778a118-6faa-4c7f-9632-b763a371ed97
Language useden
Citeable Identifier1246.145.2177
Revision Number0.0.1-alpha
protocol
ModalityModality:
  • unmasked air conduction
  • masked air conduction
  • unmasked bone conduction
  • masked bone conduction
  • soundfield
  • other
Tested SideTested Side:
  • Left ear [Left ear was tested]
  • Right ear [Right ear was tested]
  • Binaural [Binaurally tested]
Signal SourcesSignal Sources: Stimulus properties (e.g., test or masking stimulus).
If masked threshold is measured, this cluster occures twice in template for stimulus and masker.
DeviceDevice: Description of the signal source used for stimulation
  • Air conduction Headphones
  • Bone conduction headphones
  • Loudspeaker
StimulusStimulus: Identification of the stimulus used in the hearing test to measure the hearing threshold.
  • Warble Tone [The test stimulus is a frequency modulated tone.]
  • Pure Tone [The test stimulus is a pure tone.]
  • Narrow Band Noise [The test stimulus is a narrow band noise centred on the specified frequency.]
PositionPosition:
  • Left ear
  • Right ear
  • Forehead
CommentComment: Additional narrative about the test results and intepretation not captured in other fields.
Hearing DevicesHearing Devices: Hearing devices active during measurement.
In template: repeat for all hearing devices of the patient (activated during measurement).
Device TypeDevice Type:
  • Hearing Aid
  • Bone Anchored Hearing Aid
  • Middle Ear Implant
  • Cochlea Implant
  • Auditory Brainstem Implant
  • Other
PositionPosition:
  • Left ear
  • Right ear
CommentComment: Additional narrative about the test results and intepretation not captured in other fields.
E.g., programs/fittings of HA or CI used during measurement.
Test environmentTest environment: The environment in which the audiometric test is administered.
This data element is commonly used in industry hearing programs.
  • Audiometric booth [Sound-treated test environment that meets audiometric standards for ambient noise.]
  • Non-sound treated room [Test environment that does not meet audiometric standards for ambient noise.]
Calibration reference dBCalibration reference dB: Scale used for acoustic calibration of the test signal.
  • dB SPL [The test stimuli are calibrated using the sound pressure level scale.]
  • dB HL [The test stimuli are calibrated using the hearing level scale.]
  • dB nHL [The test stimuli are calibrated using the normal hearing level scale.]
CommentComment: Additional narrative about the protocol for the audiogram not captured in other fields.
For example, the co-operation of the test subject or description of the method used for testing.
data
Spectrum
FrequencyFrequency: The stimulus frequency tested.
Frequency is not applicable if a click stimulus is presented.
Property: Frequency
Units:
  • 125.0..16000.0 Hz
    Limit decimal places: 0
  • 0.125..16.0 kHz
    Limit decimal places: 3
Signal LevelSignal Level: The stimulus intensity at threshold for the test frequency.
Property: Loudness
Units: -10.0..120.0 dB
Limit decimal places: 0
ResponseResponse:
  • No perception [Maximum level presented without obtaining a threshold]
  • Detection threshold
  • UCL [Uncomfortable Loudness Level]
  • Non-auditory perception [E.g., vibration]
  • Contralateral perception [Stimulus is perceived on the non-tested side]
Masking LevelMasking Level: The stimulus intensity at threshold for the test frequency.
Property: Loudness
Units: -10.0..120.0 dB
Limit decimal places: 0

Possible reasons why null:
  • unknown
  • masked
  • not applicable
  • no information
events
Any eventAny event: @ internal @
Other contributors
Translators